3. Bonding: Apply a uniform coat of Peak
with an Inspiral
Brush tip. Lightly brush for 10 seconds. Using air syringe,
®
gently thin adhesive to a uniform glossy coat and continue drying for 10 seconds.
Avoid pooling of adhesive. Surface will look glossy. Light cure for 20 seconds
using a standard light with an output less than 1000mW. Light cure 10 seconds
using a light with an output greater than 1000mW.
4. Restore: Build incrementally with an appropriate composite following
manufacturer's directions.
Precautions:
1. Resins can be sensitizing. Dental personnel should avoid repeated contact of
uncured dental resin with skin. Do not use on patients with a known sensitivity to
acrylates or other resins.
2. Remember that light activated bonding agents are sensitive to ambient light;
syringe luer lock caps should be replaced following use to prevent unwanted
polymerization. We recommend covering syringe tip with 2 x 2 gauze if left
exposed to ambient light for long periods of time during a procedure.
3. Resins should be refrigerated for long-term storage to maintain shelf life.
4. For chemical cure or dual cure luting, use the PermaFlo
5. When dispensing intra-orally, never force syringe plungers. If there is excess
resistance to flow, replace tip and proceed.
6. Use only recommended tips with each material.
7. Tips should not be re-used so to prevent cross-contamination.
8. To optimize bond strength, use oil-free, moisture-free air.
LC Bond Resin onto etched surface
®
DC system.
®