Calibration Mode
GB
Store
Date
RESULT
Time
Cal
Zero
Mode
Test
On
Off
SEL
CAL
SEL
cAL
Mode
SEL
CAL
On
Off
MiniDirect_Cl_1 09/2008
Factory calibration reset
Resetting the user calibration to the original factory calibration will
reset all methods and ranges .
A user calibrated method is indicated by an arrow while the test result
is displayed .
To reset the calibration press both the [MODE] and [ZERO/TEST] key
and hold .
Switch the unit on using the [ON/OFF] key .
Release the [MODE] and [ZERO/TEST] keys after approx . 1 second .
The following messages will appear in turn on the display:
The factory setting is active .
(SEL stands for Select)
or:
Calibration has been set by the user .
(If the user calibration is to be retained, switch the unit off using the
[ON/OFF] key) .
Calibration is reset to the factory setting by pressing the [MODE] key .
The following messages will appear in turn on the display:
Switch the unit off using the [ON/OFF] key .