A
1.25
µS/cm
A
25.00
°C
Cal Compensation
Standard
H
40.5
mS/cm
A
23.9
°C
Conductivity Calibration
Type = Zero Point
A
1.035
mS/cm
A
21.9
°C
A Point1 = 0.000 mS/cm
A
C = 1.035 mS/cm
Cond Ind One point Slope Calibration
217.4
µS/cm
A
25.0
°C
Conductivity Calibration
Type = 1 Point Slope
217.4
µS/cm
A
25.0
°C
A Point1 = 215.0 µS/cm
A
C = 217.4 µS/cm
16
Select the kind of calibration (standard, linear or n. water).
Select Zero point and press [ENTER]. Remove the sensor and dry it.
Press the [ENTER] key again to go on with the calibration.
As soon as the conditions have stabilized (or [ENTER] pressed in the manual mode)
the display changes to show the multiplier and adder.
After a successful calibration, the calibration values are taken over and stored in the
cal history (Adjust), only stored in the cal history (Calibrate) or the calibration can be
aborted (Abort).
Start as in chapter Cond Ind Zero point Calibration described. Select the kind of calib-
ration.
Select 1 point Slope and press [ENTER].
Enter the value of calibration point 1 and then press the [ENTER] key to start calibra-
tion.
The cell multiplier and the adder are displayed. After a successful calibration, the
calibration values are taken over and can be stored in the cal history (Adjust), only
stored in the cal history (Calibrate) or the calibration can be aborted (Abort).