7. Release the buttons
a. Below is the information that will be shown:
RATE Display
VOLUME DELIVERED Display
4r
405
8. If pump RATE and VOLUME DELIVERED displays
match the information shown in step 7.a., the
pump has the Free Flow Alarm feature installed
9. Press and hold the OFF key to exit the Free
Flow Alarm test mode
Symbols
Serial Numbers less than 700000
Degree of Protection
IPX1
Provided by Enclosure
Medical Electrical Equipment
1. Classified with respect
to electrical shock, fire
and mechanical hazards in
accordance with UL60601-1
2. Classified with respect
to electrical shock, fire,
mechanical and other specified
hazards in accordance with
CAN/CSA C22.2 No.601.1
Serial Numbers 850000 or greater
Class II Equipment
IP21
Degree of Protection
Provided by Enclosure
Follow Instructions
For Use
Alternating Current
ON (Power)
OFF (Power)
Intertek ETL Classified for Medical
Electrical Equipment to IEC 60601-
1:2005, IEC 60601-1-11: 2010, CSA
C22.2#60601-1:2008
All Serial Numbers
Type BF Applied Part
(Administration Sets)
Year of Manufacture
Attention: Consult
Accompanying Documents
Transportation/Storage
Temperature Limits
Transportation/Storage
Humidity Limits
Manufacturer
Non-Ionizing
Electromagnetic Radiation

Controls

On, Applies power to unit.
1
Off, Turns power off except to battery charger.
2
Inc, Increases flow/infusion rate when pressed
3
and held. Increases dose rate after Dose Limit
Set Key has been pressed.
Dec, Decreases flow/infusion rate when pressed
4
and held. Decreases dose after Dose Limit Set
Key has been pressed.
Run/Hold, Alternately starts and stops unit.
5
Corrects alarm condition and restarts unit when
pressed twice.
Total Reset, Resets volume delivered display to 0.
6
Dose Limit Set, Allows dose to be set and
7
displays the dose limit.
Infusion Rate Display, Indicates delivery rate.
8
Volume Delivered Display, Constantly
9
EN
3