EN
6
Situation
Battery light is on
LOW BAT alarm activated
LOW BAT light is on
OCCL/EMPTY
alarm activated
Troubleshooting Guide
Possible
Causee
Pump is operating on battery power
Battery charge is below performance level
Battery life is fifteen minutes or less
Feeding container is empty
Occlusion is restricting flow
Set drip chamber not properly placed in pump
Pump is not in an upright/vertical - position or not clamped
properly to the pole
Pump is on hold for more than 2½ minutes**
Bright light is reaching drop sensor
Drip chamber walls are coated with formula or condensation
Drop sensor lenses in drip chamber guide are dirty
Drip chamber is overfilled
Solution
Plug cord into AC outlet
Plug cord into AC outlet
Plug cord into AC outlet
To avoid degrading the battery life keep battery charged
If low battery persists then return pump for service
Replace or refill feeding container
Identify point of occlusion
- Pump set
•kinked tubing
•blocked tubing
•closed roller clamp
- Feeding tube
- Feeding container
Assemble properly
Straighten the pump and clamp properly to the pole
Press RUN/HOLD button
Turn pump so sensor is not in direct light
Manipulate chamber to clear drops in sensor path
Clean with cotton swab dampened with isopropyl alcohol
Replace with new Administration Set