Extraoral Suction System, Instructions For Use and Installation

7.7 Fuse Replacement

•Turn off power switch, remove power cord.
•Keep your hands dry.
① Remove power cord from wall.
② Open the cradle cover using a Philips
screwdriver as instructed on fuse cradle.
③ Replace with a new fuse (6GFU-F15A250V)
④ Reinstall the cradle cover.
⑤ Plug in the power cable.
⑥ Turn on the power to continue operation.

8 Daily Maintenance

To ensure a clean daily use, the mouthpiece hood and clean frequently.
•To disinfect suction tubes, use a high alcohol content disinfectant spray.
•To disinfect suction hood, avoid using disinfectant liquid that may change the shape or
color of the hood.

8.1 8.1Maintenance Period of Equipment Parts

Frequency
Before Each Business Day
On a Per Patient Basis
After Each Business Day
Every 6 Months
Every 12 Months
Every 12 Months
Content
Check equipment surfaces and parts for any irregularities
Suction hood Disinfection and Arms for interior Disinfection
Clean the equipment
Fine filter replacement
HEPA Filter Replacement
Noise filter stick replacement
20