Cleaning The Charger - Medtronic Guardian 2 Link Manual De Instrucciones

Tabla de contenido
MP6025871-12C1 / A
2
Holding the tester, saturate a cotton swab in the Detachol solution and gently rub
the adhesive residue on the transmitter until it is fully removed.
3
Continue with the cleaning procedure. See Cleaning the transmitter, on page 9
for details.

Cleaning the charger

This procedure is for general cleaning as required, based on physical appearance.
Caution:
Caution:
To clean the charger:
1
Wash your hands thoroughly.
2
Use a damp cloth with mild cleaning solution, such as a dishwashing detergent,
to clean any dirt or foreign material from the outside of the charger. Never use
organic solvents, such as paint thinner or acetone, to clean the charger.
3
Place the charger on a clean, dry cloth and air dry for two to three minutes.
Troubleshooting
The following table contains troubleshooting information for the transmitter, charger,
and tester. For more information about troubleshooting, see your pump user guide.
Problem
You connected the
transmitter to the
charger and no lights
came on.
The charger is NOT waterproof. Do NOT immerse in water or any
other cleaning agent.
Dispose the charger according to the local regulations for battery
disposal (non-incineration).
Likely Cause(s)
The transmitter connec-
tor pins are damaged or
corroded.
Your charger battery has
no power.
Resolution
1
Check the transmitter connector pins for
damage or corrosion. For more informa-
tion about your connector pins, see In-
specting the transmitter connector pins,
on page 8. If the pins are damaged or cor-
roded, contact the 24 Hour HelpLine or
your local representative. It may be time
to replace your transmitter.
2
If there is no damage to the connector
pins, replace the battery in the charger.
For instructions on replacing your charger
battery, see Installing a battery in the
charger, on page 4.
-11-