Instructions
Suspending Drug Therapy - Pump Rinse
10. Repeat Steps 4 through 9 for the second pump rinse.
11. After the second rinse, repeat Steps 6 and 7 to refill the pump reservoir with 20 mL sterile saline
solution.
12. When the pump refill is completed, clamp the extension tubing and remove the needle from the refill
septum. Discard the tubing, syringe and the needle.
13. Program the refill using the Prometra
Calculations
Please refer to the appropriate supplementary Calculations Guide.
Trademarks are the property of their respective owners.
US and Foreign patents issued and pending. Please consult flowonix.com for the most up-to-date
information.
Rx Only
REFILL KIT, For use with Prometra
Pump Programmer.
®
Programmable Infusion Systems
®
Page 9