6 Data logger

The instrument has two different logger functions for recording data:
• "Func-Stor": manual recording of measurement values by button press. A measuring
location entry (L ID) is also required.
• "Func-CYCL": automatic recording of measurement values at intervals of the set
cycle time.
The logger records 2 measurement results for each record. A record consists of:
• Measurement value pH or mV/mVH
• Measurement value of the temperature (°C / °F)
• Measuring location L ID (only with "Func-Stor")
• Time and date at the time of saving
i
The real-time clock is required for the chronological assignment of logger data
and calibration times. The settings under <SEt InSt> should therefore be
checked if required.

6.1. Func-Stor

Manual recording of measurement values
In this mode, up to 1000 records can be saved. To do this, select the item Func = Stor in
the configuration menu <SEt LoGG>. An indicator arrow ( ) now appears on the display
above logg. From now on, measurement data can be stored in operating mode as follows:
/
i
SD_305_pH/ORP_1 04/2021
Briefly pressing the <Store> button stores a record in operating mode
Select a measurement ID from "L ID" (0 ... 9999). This function allows
you to assign the measured value to a sample or measuring location.
Confirm the entry
If the logger memory is full, the following message appears
on the display: Data are retrieved and deleted via the data
archive.