Description Of Functions - Tintometer Lovibond SD 305 pH/ORP Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 87

4.1.2 Description of functions

Menu
Parameter
Set parameter: Setting of the measuring parameter
40
Values
Meaning
Selection of the displayed resolution
0.01
e.g. pH 7.01
0.001
e.g. pH 7.001
Calibration: Selection of the number
of calibration points
1 pt
1-point calibration (offset calibration,
slope - 59.2 mV/pH)
2 pt
2-point calibration (neutral & one
further buffer)
3 pt
3-point calibration (neutral & two
further buffers)
Calibration: Selection of the buffer series
Std
Standard buffer solutions
pH 4.01, pH 7.00, pH 10.01
din
Buffer solutions compliant with
DIN 19266
pH 1.68, pH 4.01, pH 6.86, pH 9.18,
pH 12.45
Edit
any buffer, manual pH value entry
Calibration:
Time interval for calibration reminder
1 ... 365
Time interval for calibration reminder
(in days)
oFF
No calibration reminder
Selection of the reference for representation
of redox measurement value (ORP)
mV
ORP vs. Ag/AgCl (3 M KCl)
mV
ORP vs. SHE
H
Selection of the temperature unit
°C
Temperature information in degrees
Celsius
°F
Temperature information in degrees
Fahrenheit
SD_305_pH/ORP_1 04/2021
*
*
Tabla de contenido
loading

Tabla de contenido