Specifications; Troubleshooting - VISIOMED TensioFlash KD-558 Manual Del Usuario

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24

SPECIFICATIONS

Product name
Model
Measuring Method
Classification
Indication
Measuring range
Accuracy
Memory
Power source
Battery life
Working Temp. & humidity
Storage Temp. & humidity
Outside dimensions
weight
Accessories
Specifications may be changed without notice in the event
of improvement being made
28
Visiomed - KD-558 - 032016
AUTOMATIC BLOOD PRESSURE MONITOR
ARMBAND
KD-558
Oscillometric System
Classe IIa
Digital LCD Display
Cuff pressure: 0~300 mmHg.
Systolic : 60~260 mmHg
Diastolic : 40~199 mmHg
Pulse: 40~180 Beat/min.
Pressure +/- 3 mmHg 
Pulse +/- 5%
2x30 measurements stored in 2 memory
banks
1.5V "AA" x 4 Alkaline Battery
Approx 500 measurements
10~40°C (50~104°F). ≤90% RH
-20~55°F (-4~131°F). ≤90% RH
Pression : 80KPa-105KPa
138 mm (La)/ 98 mm (Lo)/ 48 mm (H)
(5 7/16" X3 27/32" X1 7/8")
Approx 211 g (Apparatus and Battery)
(7 7/16 oz.)
Cuff: fit arm 22~32 cm (8,7 -
12,7 in) 30~42cm (11 13/16" - 16 17/32")
(optional) 42~48cm (16 17/32 "- 18 29/32")
(optional). User manual, carrying case, 4
batteries 1.5V AA

TROUBLESHOOTING

If you have one of the following problems, please refer to
this breakdown service guide to help resolve the problem. If
the problem persists, please contact our customer service at
+33 892 350 334 (0,34€/mn).
PROBLEM
POSSIBLE CAUSE
The cuff position was not
correct or it was not properly
tightened
Body posture was not correct
during testing
LCD Display shows abnormal
result
Speaking, arm or body
movement, angry, excited or
nervous during testing
Irregular heartbeat
(arrhythmia)
LCD shows low battery
Low Battery
symbol
Pressure system is unstable
LCD shows "Er 0"
before measurement
Fail to detect systolic
LCD shows "Er 1"
pressure
Fail to detect diastolic
LCD shows "Er 2"
pressure
Pneumatic system blocked
LCD shows "Er 3"
or cuff is too tight during
inflation
Pneumatic system leakage
LCD shows "Er 4"
or cuff is too loose during
inflation
Visiomed - KD-558 - 032016
SOLUTION
Apply the cuff correctly and
try again
Review the "BODY POSTURE
DURING MEASUREMENT"
sections of the instructions
and re-test.
Re-test when calm and
without speaking or moving
during the test
It is inappropriate for people
with serious arrhythmia
to use this Electronic
Sphygmomanometer.
Change the batteries
Don't move and try again.
Apply the cuff correctly and
try again
29
loading