Care and Maintenance
12
Installationsanleitung CAIXS2
Cleaning the Stainless Steel Surfaces
–
Only use conventional household cleaning agents that are suitable for
stainless steel.
–
Only use solvents for cleaning stainless steel parts.
All stainless steel parts should be cleaned at regular intervals: Rub stainless
t
steel surfaces with a moist cloth, with a cleaning agent if required, then
remove all residue from the surface.
Allow device to dry. For additional protection, protective oil may be applied.
t
Replacing the Dust Cover
Damaged dust covers should be replaced immediately.
Remove damaged dust cover.
t
Place the new dust cover on the indicator and press it over the edge of the
t
front and rear side of the device until it is fixed in place.
Safety Inspections
Safe operation of the device is no longer ensured when:
–
there is visible damage to the device or power cord
–
the integrated power supply for the indicator no longer functions properly
–
the device has been stored for a relatively long period under unfavorable
conditions (e.g. excessive humidity)
If there is any indication that safe operation of the device is no longer
warranted:
Disconnect the power supply to the device (unplug the power cord from the
t
mains supply) and make sure the device cannot be used for the time being.
Notify your nearest Minebea Intec Service Center.
t
Maintenance and repair work may only be carried out by service technicians:
–
who have access to the required maintenance manuals
and
–
who have attended the appropriate training workshops.
The seals on the device indicate that the device may only be opened and
h
maintained by authorized specialist personnel, so that the correct and safe
operation of the device is ensured and the guarantee remains valid.