Stainless steel or
Hastelloy pipe
connection
PVDF pipe
connection
We have reviewed the contents of this publication to ensure consistency with the hardware and software described. Since variance cannot be precluded entirely, we cannot guarantee full con-
sistency. However, the information in this publication is reviewed regularly and any necessary corrections are included in subsequent editions. Responsibility for suitability and intended use of
this instrument rests solely with the user.
Copyright © Siemens AG 05.2008 All Rights Reserved. Technical data subject to change without prior notice.
Siemens A/S - Flow Instruments
6430 NORDBORG
DENMARK
www.siemens.com/flowdocumentation
Order no.: 083R9050
Literature no.: SFIDK.PI.027.U5.83
Assembling (EPDM, PTFE and Graphite gaskets)
The gaskets are placed in the sensor. Be aware that the gaskets are centrered. The pipe connection
to be fastened with screws 1 and 2, which are to be tightened loosely. Hereafter the last four
screws are assembled.
The screws to be "cross-tightened" (1, 2, 3, 4, 5, 6) to 3 Nm.
For PTFE gaskets use M4 x 14 screws.
Assembly (PTFE gasket)
Place one gasket in the sensor, then position the earthing ring and another gasket, and finally the
PVDF pipe connection and tightening plate. Make sure the gaskets and earthing ring are centered.
Fasten the pipe connections using screws 1 and 2 (M4 x 25), without tightening fully. Then fit the
last four screws. Position the third gasket on the sensor. Make sure the gasket is centered. Fasten
the pipe connection and tightening plate with screws 1 and 2 (M4 x 20), without tightening fully.
Then fit the last four screws.
The screws to be "cross-tightened" (1, 2, 3, 4, 5, 6) to 3 Nm.
*083R9050-05*