3. Data memory
To display the last temperature measurement, press the MEM button.
You will obtain the last temperature measurement. To switch on the last measurement, press
the °C/°F button.
The number indicated in interval of two measures corresponds to the number of measurements.
Press on the button MODE for review of the previous measurements.
To exit data memory, press the measure key. The Thermoflash® LX-260T EVOLUTION switches
off after 5 seconds.
4. Choosing the language
Use the Language key to choose a language for the vocal function. The choice of language is
visible.
Languages available: French, English, Spanish, Italian, German, Dutch, Silent Function.
By default the ThermoFlash® LX-260T EVOLUTION is in French.
5. Alarm setup
The alarm threshold default value is 38°C (100.4°F). It cannot be changed. In case of higher tem-
peratures than 38°C, a signal beeps while measuring the temperature.
6. Changing the batteries
Display: when the LCD screen displays Battery, the battery is used.
Operation: Open the lid and change the batteries, taking great care with the correct positioning.
A mistake with this could cause damage to the apparatus and compromise the guarantee of
your ThermoFlash® LX-260T EVOLUTION. Never use rechargeable batteries. Use only batteries
for single usage.
VIII. TECHNICAL CHARACTERISTICS & PRECISION
1. Normal conditions of use operating temperature: 10°C ~ 40°C (50°~ 104°F). Normal conditions
of storage and transportation temperature: -20°C ~ 50°C (68°~ 122°F). Humidity rate: 10 - 85%
2. Power: DC3 V (2 batteries LR6).
3. Size: 220 x 145 x 48 mm.
4. Weight: 180g.
5. Display Resolution: 0.1° C (0.1°F).
6. Measuring range:
In BODY mode: 32°C – 42.9°C (90°F – 109°F).
In SURFACE TEMP mode: 0°C – 60°C (32°F to 140°F).
7. Precision: From 36°C to 39°C (96.8°F to 102.2°F) = +/-0.2°C.
8. Consumption: ≤ 50mW.
9. Measuring distance: 2-5cm.
10. Automatic stop: 5 sec.
22
Visiomed - LX-260TE - 032017
Longevity use
The ThermoFlash® LX-260T EVOLUTION is guaranteed for 40.000 readings.
IX. ADVICE
- The protective glass over the lens is the most important and fragile part of the thermometer,
please take great care of it.
- Clean the glass with cotton fabric, wet with water or 70° alcohol.
- Do not use other batteries than mentioned batteries, do not recharge non rechargeable bat-
teries, do not throw in fire.
- Remove the batteries when thermometer is not used for an extended period of time.
- Do not expose the thermometer to sunlight or water.
- An impact will damage the product.
X. ACCESSORIES SUPPLIED
User manual, Guarantee card, Batteries supplied (LR6).
XI. TROUBLESHOOTING
If you have one of the following problems while using your ThermoFlash® LX-260T EVOLUTION,
please refer to this breakdown service guide to help resolve the problem. If the problem persists,
please contact our customer service at +33 892 350 334.
- The screen displays temperature superior to 95°: the temperature is in Fahrenheit. Change
the measurement unit to Celsius.
- The screen displays the body temperature inferior to 32°C (89.6°F): to measure a body
temperature, the BODY mode must be selected. If you are on SURFACE TEMP mode, the 32°C
(89.6°F) temperature displayed is showing the external temperature that your body releases.
- The screen displays the message HI: when using the ThermoFlash® LX-260T EVOLUTION,
the message HI can show on the screen. The analysis is above the measurement range selected,
either superior to 42.9°C (109°F) in BODY Mode or superior to 60°C (140°F) in SURFACE TEMP
Mode.
- The screen displays the message LO: when using the ThermoFlash® LX-260T EVOLUTION,
the message Lo can show on the screen. The temperature analysed is under the measuring
range selected, either less than 32°C (90°F) in BODY Mode or less than 0°C (32°F) in SURFACE
TEMP Mode.
Visiomed - LX-260TE - 032017
23