Trouble Shootings - Gima RS01 Manual De Usuario

Ocultar thumbs Ver también para RS01:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

9 Trouble Shootings

Trouble
SpO
or Pulse rate is
2
displayed abnormally.
Nose air flow waveform
display is instable.
Powering on disabled
PC can not identify SD
card disk symbol.
In full power,the opera-
tion time is too short or
the charging time is al-
ready over 10 hours, the
battery is still not full.
The device may not meet your intended purpose owing to operation or other prob-
lems, and consult the following troubleshooting to solve the problems you have met.
Appendix 1
State
Low voltage alarm
SpO
alarm
2
Pulse rate alarm
Probe error alarm
ENGLISH
Possible cause
1.The finger is not put in cor-
rectly.
2.Patient SpO
2
the measurement range.
3.The deepness of the finger
putting in is not enough.
4.The finger wobbles or the
patient is in movement.
1.Connection problem.
2.Wearing method problem.
3.The patient is in movement.
1.Battery capacity lack or no
power.
2.Device failure.
1.The USB connection has
been worn and in bad con-
nection.
2.SD card is damaged.
The battery is damaged.
Alarm condition delay
1s
330ms
330ms
16ms
Solution
1.Try again with the finger
value is beyond
2.Try repeatedly,if you are
3.Try again with the finger
4.Please keep calm.
1.Correctly connect.
2.With
3.The patient shall keep
1.Charge the battery at
2.Contact your local ser-
1.Try to connect the USB
2.Contact your local ser-
Contact your local service
center
Alarm signal generation delay
20ms
20ms
20ms
20ms
correctly put in.
sure the product perfor-
mance is normal,go to
hospital for diagnose.
correctly putting in.
correct
wearing
method.
calm.
once.
vice center
data line repeatedly.
vice center.
40
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

33676

Tabla de contenido