41
Appendix 2
Guidance and manufacturer's declaration – electromagnetic emissions – for
all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emission
The RS01 is intended for use in the electromagnetic environment specified below. The
customer of the user of the RS01 should assure that it is used in such and environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Guidance and manufacturer's declaration – electromagnetic immunity – for
all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The RS01 is intended for use in the electromagnetic environment specified below. The
customer or the user of RS01 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
Compliance
Electromagnetic environment – guidance
Group 1
The RS01 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
Class B
The RS01 is suitable for use in all establishments
other than domestic and those directly connected
to a low voltage power supply network which sup-
plies buildings used for domestic purposes.
IEC 60601
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3A/m
Class B
ENGLISH
Electromagnetic
environment – guidance
Floors should be wood, concrete
or ceramic tile. If floor are cov-
ered with synthetic material, the
relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.