11.7. Validation of sensor measurements – vacuum gauge
For the PROCOM-2 / CYCLIC 2020 and MILLENIUM:
connect the electronic calibrating vacuum gauge to the bleed valve of the receiver and check the sensor
regulation and the mechanical vacuum gage
11.8. Checking operation of the network alarm
Check with relevant authority that this test is suitable for the activity and architecture of
the building in which the network is installed. This action will trigger an alarm.
Disconnect the 'rilsan' pipe from the network alarm sensor C2V to make a vacuum drop; the low network
alarm will be triggered at –370 mbar.
PRECAUTIONS WITH THE NETWORK
The components of the vacuum unit are not designed to resist an overpressure coming from the network,
for example during the cleaning and pressure tests phases.
It is the same for the lowest level drip pot placed at the various points of the network.
It is imperative to be isolated from the network during maintenance operations.
The suction through network of hydrogen peroxide or all other having the same properties
is completely proscribed.
The hydrogen peroxide can release from oxygen and under certain conditions, to cause an increase
abnormally high of the pressure.
TRAINING
In order to ensure perfect aptitude in the use of the vacuum plant for medical use, the MIL'S SAS Company
offers its own training for the use of this equipment. It is intended for users and technical personnel working
on or near the vacuum plant for medical use.
CLEANING
14.1. Units cleaning
Do not use corrosive product
If the units system requires cleaning, use a soap solution and a clean cloth.
Then wipe well with a dry cloth.
For disinfection, use swabs soaked in a disinfectant solution.
Frequency: as required.
14.2. Network cleaning
When cleaning a network with a cleaning solution or disinfectant, we must be vigilant.
The solution mixed with the organic material present in the network, could lead to an extremely rapid and
significant pressure for progress in specific components (jars low point, for example) which can lead to an
explosion of them.
520713-00
MINIVAC 1/2/3 PS & ARICA
ELECTROMAGNETIC COMPATIBILITY
See Annex
18