Guidance and Manufacture's declaration – electromagnetic immunity-
For all EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and Manufacture's declaration - electromagnetic immunity
Guidance and Manufacture's declaration – electromagnetic environment specified below. The customer or
the user of the Pulse Oximeter (18710) should assure that it is used in such an environment.
IEC 60601
Immunity
test level
test
3 V/m
Radiated
80 MHz to
RF
2.5 GHz
IEC
61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations, Electromagnetic propagation is affected by
absorption and reflection structures, objects and people.
Compliance
Level
Portable and mobile RF communications equipment should be
used no closer to any part of the Pulse Oximeter (18711), including
cables, than the recommended separation distance calculated
3 V/m
from the equation applicable to the frequency of the transmitter.
Recommended separation distance
3.5
d=
E
1
7
d=
E
1
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey a , should be less than the compliance
level in each frequency range. b
Interference may occur in the vicinity of equipment marked with
following symbol:
Electromagnetic Environment – guidance
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz
P
- 12 -