Drive Health Ox 18710 Manual Del Usuario página 13

Oxímetro de pulso tactil para dedo
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

14. Possible Problems and resol
Problems
SpO
% or pulse
1. Finger is not plugged correctly
2
rate can not be
2. User's SpO
shown normally
1. Finger might not be plugged deep
SpO
% or pulse rate
2
is shown unstably
2. Excessive user movement
1.No battery or low power of battery
The Monitor can not
2.Battery might be installed incorrectly
be powered on
3.The Monitor might be damaged
1.The Oximeter is automatically
Indication is
suddenly off
2.The batteries power is too low to work
1.Err 3 means the red emission LED is
"Error3" or "Error4"
Displayed on screen
2.Err 4 means the infra-red emission
Error 6
Err 6 means the crystal is failure
Err 7 means all the emission LED or
"Error7" displayed
reception dioxide is damaged.
on screen
Possible reason
value is too low to be
2
measured
enough
powered off when no signal is
detected longer than 8 seconds
damaged.
LED is damaged.
1. Retry by plugging the finger
2. There is excessive illumination
3. Measure more times, If you can make
sure about no problem existing in the
product. Please go to a hospital timely
for exact diagnosis
1. Retry by plugging the finger
2. Be calm.
1. Please replace battery
2. Please reinstall the battery
3. Please contact with local customer service centre
1. Normal
2. Replace the battery
1. Check the red emission LED
2. Check the infra-red emission LED
Change the crystal
Check the emission LED and reception
dioxide.
- 09 -
Solution
Tabla de contenido
loading

Tabla de contenido