Drive Health Ox 18710 Manual Del Usuario página 15

Oxímetro de pulso tactil para dedo
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Guidance and Manufacture's declaration – electromagnetic immunity-
Guidance and Manufacture's declaration - electromagnetic immunity
Guidance and Manufacture's declaration – electromagnetic environment specified below. The customer or
the user of the Pulse Oximeter (18710) should assure that it is used in such an environment.
Immunity test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
For all EQUIPMENT and SYSTEMS
IEC 60601 test level
Compliance Level Electromagnetic Environment – guidance
+/- 6kV contact
+/- 8kV air
3A/m
Floors should be wood, concrete
or ceramic tile. If floor are covered
+/- 6kV contact
with synthetic material, the relative
+/- 8kV air
humidity should be at least 30%.
Power frequency magnetic fields
should be at levels characteristics of a
3A/m
typical location in a typical commercial
or hospital environment.
- 11 -
Tabla de contenido
loading

Tabla de contenido