Instructions for use

Cancelling a reading

When a SCBI is removed from it's position during incubation
process, an audible alarm will sound to indicate the tube should be
returned to its corresponding position.
If the indicator is not returned to the reading position within 10
seconds from its removal, the reading will be canceled and the auto-
reader will print out a ticket confirming the cancellation.
In the hygiene monitoring system, the user can cancel the reading
by pressing the button
of the chosen position and the button
for three seconds. The device will cancel the reading and the
auto-reader will print out a ticket confirming the cancellation.

Temperature monitoring

The Bionova
IC10/20FRLCD Auto-reader contains an internal
®
temperature control. If the temperature falls outside of the
specified range (37±2 or 60±2 ºC), the blue temperature led light
will start to twinkle.
Temperature can be externally monitored by using a Bionova
IC1020 thermometer or other similar. IC10/20FRLCD Auto-reader
has a special hole for placing the thermometer which is located at
the back of the device.

Thermal paper specifications

Recommended paper: JUJO AF50KSE3 or similar (order code
ICTP).
Paper width : 58 mm
Maximum paper thickness: 60 g/m
Maximum diameter size: 50 mm

Replacement of the paper roll

The printer uses thermal paper rolls.
TB-
®
2
1.Pull the handle of the
printer's door.
2.Open the printer's cover
and remove the spent roll.
IC10/20FRLCD Page 10
Remaining incubation time |
Configured programs
Bionova
IC10/20FRLCD Auto-reader allows to verify remaining
®
time of incubation and the configured programs of each reading
position in a very simple way. To do this, press on the icon
Then, a new screen will open indicating the remaining readout
time of each position that is performing an incubation. In
positions where there is no readout initiated, the name of the
selected program will be displayed.
Press
to return to the main screen or just wait one minute to
return automatically.

Time setting mode

Modification of the time using the Bioupdate
Connect the auto-reader to the same network as your PC using
a specific Ethernet cable and start the program. Select the
Bionova
IC10/20FRLCD Auto-reader and press the Clock Sync
®
button, the auto-reader will be synchronized with the PC's clock.
Synchronization can only be done as long as there is not an
ongoing reading.
Change the time using the Bionova
Connect the auto-reader to the same network as your PC using
a specific Ethernet cable and start the program. The Bionova
3.Place the new paper roll
with the outer side up.
4.Close the printer cover by
pressing on the sides of the
lid.
.