ELECTROMAGNETIC COMPATIBILITY

AViTA NT13 is intended for use in the electromagnetic environment specified below. The
customer or the user of AViTA NT13 must make sure that it is used in such an environment.
Guidance and manufacturer's declaration - Electromagnetic emissions
Phenomenon
Conducted and radiated
RF EMISSIONS
Harmonic distortion
Voltage fluctuations
and flicker-ing
a) The equipment is suitable for use in Home Health Environments and Professional Health
Care Environments limited to patient rooms and respiratory treatment facilities in hospital
or clinics. The more restrictive acceptance limits of Group 1 Class B (CISPR 11) have been
considered and applied. The equipment is suitable for use in the mentioned envi-ronments
when directly connected to the Public Mains Network.
b) The test is not applicable in this environment unless the ME EQUIP-MENT and ME SYSTEM
used will be connected to the PUBLIC MAINS NETWORK and the power input is otherwise
within the scope of the Basic EMC standard.
Guidance and manufacturer's declaration - Electromagnetic immunity - Enclosure port
Phenomenon
ELECTROSTAT-IC
DISCHARGE
Radiated RF EM fields
Proximity fields
from RF wireless
communications
equipment
Professional healthcare
facility environment a)
a)
Not applicable
Not applicable
Basic EMC
standard or test
method
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
HOME HEALTHCARE ENVIRONMENT a)
CISPR 11
Group 1 Class B
Immunity test levels
Professional
healthcare facility
envi-ronment
± 8kV contact
± 2 kV, ±4kV ±, ±8 kV, ±15 kV air
a)
COMPLIANT
NOTE: Further information about distances
to be maintained between portable and
mobile RF communica-tions equipment
(transmitters) and the AViTA NT13 can be
requested from AViTA using the contact infor-
mation provided in this manual. However, it
is advisable to keep the electromechani-cal
aerosol equipment at an ade-quate distance
of, at least, 0.5 m from mobile phones or
other RF communications transmitters to
mini-mise possible interference.
HOME HEALTHCARE
ENVIRONMENT
10 V/m b) 80MHz - 2.7
GHz 80% AM at 1kHz
ENGLISH • 19