ELECTROMAGNETIC COMPATIBILITY

The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare and clinic environments.
WARNINGS
• Don't near active HF surgical equipment and the RF shielded room of an ME system for
magnetic resonance imaging, where the intensity of EM disturbances is high.
• Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
• Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation."
• Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the equipment JXB-313, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE
with regard to electromagnetic disturbances for the excepted service life.
Portable and mobile RF communication equipment may affect the performance of the
JXB-313, avoiding strong electromagnetic interference when used, such as near mobile
phones, microwave ovens, etc.
2. Guidance and manufacturer's declaration -electromagnetic emissions and Immunity
Guidance and manufacturer's declaration -electromagnetic emissions
and Immunity
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3

EMC Declaration

Table 1 EMISSIONS
Compliance
Group 1
Group B
Not Applicable
Not Applicable