ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer's declaration – electromagnetic immunity - for all ME equipment and ME systems
The unit with PROCOM 2, C2020 or MILLENIUM automate is intended for use in the electromagnetic environment specified below.
The customer or the user of the unit with PROCOM 2, C2020 or MILLENIUM automate should assure that it is used in such an electromagnetic environment.
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2 : 2008
Electrical fast
±2 kV for power supply lines
transient/burst
IEC 61000-4-4 : 2012
±1 kV for input/output lines
Shock waves
±1 kV mode differential
IEC 61000-4-5 : 2014
AMD1 : 2017
Transient electrical
disturbances – 12 VDC
power supply
ISO 7637‐2
UT=0%, 0,5 cycle (0, 45,
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11 : 2004
AMD1 : 2017
UT=0%; 250/300 cycles
Conducted RF
IEC 61000-4-6 :2013
6V (discrete frequencies)
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8 : 2009
Radiated RF
IEC 61000‐4‐3 : 2006 /
AMD1 : 2007 / AMD2 : 2010
15 fréquences discrètes
(tableau 9/ CEI 60601‐1‐2)
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
(1) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the units with PROCO M2, C2020 and MILLENIUM automate
is used exceeds the applicable RF compliance level above, the units with PROCOM2, C2020 and MILLENIUM automate should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the units with PROCOM2, C2020 and MILLENIUM
automate.
(2) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
520937-00
IEC 60601 test level
Finding
±8 kV contact
±15 kV air
(Note B & Note C)
±2 kV mode common
(Note E)
0, 90, 180 & 270°
600 V
90, 135, 180, 225,
270 & 315°)
UT=0%; 1 cycle –
(Note H)
UT=70% 25/30
cycles (0°)
3V (0.15‐80MHz)
(Note F)
30 A/m
50 and 60 Hz
(Note G)
3 V/m
80 MHz to 2,7GHz
9V/m to 28 V/m
(Note D)
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
PASS
covered with synthetic material, the relative humidity should be at
least 30 %.
PASS
PASS
N/A
Mains power quality should be that of a typical commercial or
hospital environment.
PASS
PASS
N/A
No components sensible in product to magnetic field.
Portable and mobile RF communications equipment should be used
no closer to any part of the units with PROCOM 2, C2020 or
MILENIUM automate, including cables, than the recommended
separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.16 √P
d = 1.16 √P from 80 MHz to 800 MHz
d = 2.33 √P from 800 MHz to 2,7 GHz
where P is the maximum output power rating of the transmitter in
PASS
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey (1) should be less than the compliance
level in each frequency range. (2)
Interference may occur in the vicinity of equipment marked with
the following symbol:
5