Disinfection System; During Treatment - Sirona C1+ Instrucciones De Uso

Ocultar thumbs Ver también para C1+:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Sirona Dental Systems GmbH
13.4

Disinfection system

C
Start
San.
13.5

During treatment

59 06 446 D 3386
D 3386.201.01.08.02
C
S
0 m
30 s
C
12:30
Desinf.
13 Water unit, assistant element
The disinfection system adds a disinfectant to the water
(1:100) to prevent the formation of microorganisms in
the water system.
Prior to initial operation
of your treatment center, sanitation must be performed.
If, on the basis of an agreement with you, sanitation was
not performed by the service engineer following installa-
tion of your treatment center, please perform sanitation
as described in the separate instructions "Care and
Cleaning by the Practice Team".
Sanitation takes approx. 24 hours.
If during treatment Desinf. flashes on the display, top up
with 1 liter of DENTOSEPT P before beginning treatment
of the next patient:
Remove the cover behind the cuspidor and fill in 1 liter
of DENTOSEPT P.
The filler trough should always be kept clean.
Make sure that no tap water enters the fill-in trough!
DENTOSEPT P
1 carton = 6x1 liter, Order No. 33 18 156
Microbiological water test
(after a working pause of more than 1 week)
Sanitation of the unit
Perform sanitation every 4 weeks. Only in this way can
you ensure that biofilm is effectively controlled in the
water paths.
(see instructions "Care and Cleaning by the Practice
Team").
CAUTION
In Japan the disinfectant OXYDOL available there must
be used instead of DENTOSEPT P.
OXYDOL must be mixed 1:1 with distilled water before it
is filled into the disinfection system.
13.4 Disinfection system
47
Tabla de contenido
loading

Tabla de contenido