Hygiene Treatment; Cleaning; Disinfecting - Seca 651 Manual De Instrucciones

Ocultar thumbs Ver también para 651:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

8. HYGIENE TREATMENT

8.1 Cleaning

8.2 Disinfecting

6. Confirm the prompt on the display with the
You exit the menu automatically.
The device will be reset to factory settings.
WARNING!
Electric shock
The device is not de-energized when the display goes out. Use of
fluids on the device may cause electric shock.
► Disconnect the power supply connector before performing any
hygiene treatment.
► Ensure that no fluids penetrate the device.
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
► Use alcohol-based disinfectant (e.g. 70 % ethanol) exclusively.
► Do not use aggressive or abrasive cleaning agents.
► Ensure that no moisture or dust get into the sensors during
cleaning.
► Use a soft cloth dampened with mild soapsuds to clean the surfaces of
the device.
1. Use an alcohol-based disinfectant (e.g. 70 % ethanol).
2. Follow the instructions on the disinfectant.
3. Disinfect the device:
► Moisten a soft cloth with disinfectant and wipe down the device with
it.
► Comply with the intervals, see table.
Interval
Before every
measurement
After every
measurement
If required
► Sterilizing
This device may not be sterilised.
Component
Weighing platform
Weighing platform
• Multifunctional display
• Display cable
Hygiene treatment • 101
English
key.
Tabla de contenido
loading

Este manual también es adecuado para:

650

Tabla de contenido