DMG MixStar-eMotion
3. Attach a new mixing tip. Make sure the mixing tip is properly seated.
4. Secure the mixing tip by sliding the locking mechanism to locking
position.
Leave used tips on the cartridge after use. They serve as a cap.
5. I nsert the cartridge from the top into the unit. A short acoustic signal
will sound when the cartridge is seated properly.
Fresh cartridges are ready for use after the first 3 cm of impression
material have been extruded and discarded.
Loading a tray
The simple two-button or foot pedal control leave both hands free to
operate the tray. The mixing speed can be increased or decreased by
pressing the Plus button [8] or the Minus button [9], which is possible
during extrusion as well. Earlier MixStar cartridges (see white slider)
should be dispensed at level 1 or 2 only.
1. S tart mixing by pressing the Start/Stop button [5] or by operating the
foot pedal. If Timer 1 is set its countdown will start now. The end of
the working time is indicated by a long acoustic signal. If a new Mix-
Star cartridge is used discard the first 3 cm of the newly mixed mate-
rial. Carefully follow the manufacturer's instructions for the selected
material. For a clearer display all elements for which no new settings
are made (-:--) are hidden after the Start/Stop button [5] is pressed.
2. Tilt the tray slightly under the mixing tip to load it. To avoid entrap-
ment of air leave the mixing tip immersed in the impression material
while filling the tray continuously with the material.
3. After the tray has been sufficiently loaded stop dispensing by press-
ing the Start/Stop button [5] or by releasing the foot pedal. Mixing is
stopped at once and pressure on the cartridge is relieved to avoid any
further flow of the material. If Timer 2 is set its countdown will start
now. 3 acoustic signals indicate when the material can be removed
from the mouth. The unit stops dispensing automatically after 2 min-
utes but can be restarted by pressing the Start/Stop button [5].
Filling an elastomer syringe
For filling elastomer syringes we recommend using mixing speed 1.
The mixing speed can be increased by pressing the Plus button [8] or
decreased by pressing the Minus [9] button. Speed adjustments are pos-
sible even during material delivery.
1. Remove the plunger of the elastomer syringe.
2. Attach the elastomer syringe with its rear end to the mixing tip.
3. S tart mixing by pressing the Start/Stop button [5] or by operating the
foot pedal. If Timer 1 is set its countdown will start now. The end of
the working time is indicated by a long acoustic signal.
4. S top dispensing into the syringe by pressing the Start/Stop button [5]
or releasing the foot pedal. Mixing is stopped at once and pressure on
the cartridge is relieved to avoid any further flow of the material.
I f Timer 2 is set its countdown will start now. 3 acoustic signals indi-
cate when the material can be removed from the mouth.
Exchanging a cartridge
1. The cartridge can only be removed if the plungers are in cartridge
exchange position. If necessary, press the Reverse button [6] and wait
[Fig. 07 - 08]
until the plungers return to their original position. The Reverse indica-
tor [J] blinks during plunger retraction. When the plungers reach the
cartridge exchange position the Reverse indicator is no longer dis-
played.
2. Release the cartridge from the cartridge compartment by pressing
with one hand from below the openings and remove it from above.
Leave used mixing tip on the cartridge after use. It serves as a cap.
3. I nsert the new cartridge into the unit from the top (see »Preparing
the cartridge for use«, step 5). In order to use up the material in the
cartridge keep a new cartridge with attached mixing tip ready before
starting the mixing process. That way you ensure that the remain-
ing material in the old cartridge does not have to be discarded as
the working time limit would be exceeded while exchanging the car-
tridge.
Attaching a mixing tip
Attach a new mixing tip before each mixing process.
1. Slide locking mechanism into open position.
2. D iscard the old mixing tip. If the cartridge was stored for a longer peri-
od of time be sure to check the openings for possible blockings and
remove them.
3. Align the mixing tip with respective outlets on the cartridge and
attach it to the cartridge.
[Fig. 09 - 10]
4. Lock the mixing tip with the locking mechanism on the cartridge.
Attach a new mixing tip before each mixing process. The device
is unable to deliver impression material through material that has
cured in the mixing tip. If the mixing tip is not properly attached to
the cartridge material may be pressed past the mixing tip and cause
soiling.
English
[Fig. 11 - 12]
[Fig. 03]
17