c) Total Measurement
1. Press the POWER button to activate the device.
2. Select the total mode with the A.TOT button.
3. Connect the input signal with BNC connector A on the front panel.
4. Put A.ATTN in the desired position. Press A.ATTN is the input signal > 300mV: it reduces the number of mistakes
since the triggering sensitivity of the input section is attenuated by a factor of 20.
5. Put the HOLD button in the depressed position and read the total from the display.
d) Check Mode
The self-check mode provides a means of verifying the overall operation of the device with the exception of the input
section, the accuracy of the time base and the time base dividers used in the period mode.
1. Press the POWER button to activate the device.
2. Select the check mode with the CHECK button.
3. Press CHECK. All segments of the display are lit in the following order: 0 → 9, "•", the LEDs of the functions
printed on the display.
4. Press the RESET button to stop the check.
7. Specifications
Display
Overrange Indication
Power Requirements
Operating Temperature
Storage Temperature
Relative Humidity
Safety
Dimensions
Weight
Frequency Measurement
Period Measurement
Time Base Error (long-term stability)
Time Base Error (short-term stability)
Max. accuracy is achieved during a one-year period after calibration. Ideal circumstances require a temperature of 18
to 28°C (64 to 82°F) and a max. relative humidity of 80%.
Channel
A
10Hz to 10MHz
100MHz
10MHz to 100MHz
B
1300MHz
100MHz to 1300MHz
DVM13MFC
Resolution and Accuracy
Frequency Measurements
Range
10MHz
100Hz, 1KHz, 10KHz
General
8 digits, 10mm high, red LEDs, with decimal point
OFL indication is lit
AC198~264V, 50/60Hz or 90~132V 50/60Hz
0°C to +40°C
-10°C to +50°C
20 to 80% RH at 0 to 40°C
in accordance with IEC1010-1 standard
300 (L) x 260 (W) x 74 (H) mm
±1850g
± 1 count ± time base error x frequency
± 1 count ± time base error x period
1 x 10
(month)
-5
5 x 10
(minute)
-6
Resolution
1, 10, 100Hz
selectable
10, 100, 1000Hz
selectable
selectable
5
Accuracy
± 1 count ± 1 x 10
x frequency (month)
-5
± 1 count ± 5 x 10
x frequency (minute)
-6