RS-232 INTERFACE

The CKT and CKT-M Series of scales include a bi-directional RS-232 interface. The scale
when connected to a printer or computer through the RS-232 interface, outputs the weight
with the selected weighing unit.
Specifications:
RS-232 output of weighing data
ASCII code
9600Baud rate(user selectable)
data bits
8
No Parity
Connector:
pin d-sub miniature socket
9
Pin 2 Output
Pin 3 Input
Pin 5 Signal Ground
The scale can be set to print text in English, French, German, Spanish, Italian or Portuguese.
The data will normally output in a label format if parameter Label=On.
4.1

INPUT COMMANDS FORMAT

The scale can be controlled with the following commands. The commands must be sent in
upper case letters, i.e. "T" not "t". Press the Enter key of the PC after each command.
Tares the scale to display the net weight. This is the same as pressing
T<cr><lf>
[Tare] ↵ key.
Sets the zero point for all subsequent weighing. The display shows zero.
Z<cr><If>
P<cr><lf>
Prints the results to a PC or printer using the RS-232 interface. It also
adds the value to the accumulation memory if the accumulation function
is not set to automatic. In CKT series, the [Print] key will either print the
current items being counted or the results of the accumulation memory
if [M+] is pressed first.
R<cr><lf>
Recall and Print- Same as if first the [MR] key and then the [Print] key
is pressed. Will display the current accumulated memory and print the
total results.
C<cr><lf>
Same as pressing [MR] first and then the [CE] key to erase the current
memory.
© Adam Equipment Company 2019