For single-patient use
Always clean the transmitter after each use.
To clean the transmitter, you need the following materials:
•  mild liquid soap
•  soft-bristled toddler toothbrush
•  container
•  clean and lint-free dry cloths
Use life
The transmitter can be cleaned up to 122 cycles or one year, whichever comes first.
Discard the transmitter at this point. If you continue to use the transmitter beyond 122
cycles or one year, the cleaning process may damage the device. Contact Medtronic to
order a new transmitter.
Warning:
Do not use the device if you see any cracking, flaking, or damage to the
housing. Cracking, flaking, or damage to the housing are signs of
deterioration. Deterioration of the housing can affect the ability to
properly clean the transmitter, and result in serious injury. Call the 24
Hour HelpLine or your local representative and discard the device
according to local regulations for battery disposal (nonincineration), or
contact your healthcare professional for disposal information.
To clean the transmitter:
 1
Wash your hands thoroughly.
 2
Attach the tester to the transmitter.
 3
If there is adhesive residue on the transmitter, see Removing adhesive residue, on
page 17.