7

Technical Data

7.1

General Data

Specification
Scale
Supply voltage
Input voltage
Power consumption
Further data
Ambient conditions
The technical specifications apply under the following ambient conditions:
Environment
Ambient temperature*
Operational capability
Storage and shipping
Relative humidity
Interface connection
Format
Parity
Transmission rate
Handshake
Electromagnetic compatibility
Interference resistance
Transient emissions
Available application programs
Power supply YEPS01-USB (optional)
USB power plug
Primary
Secondary
Further data
Unit
Value
Only via USB interface or Sartorius AC adapter YEPS01-USB
V
+5.0
DC
W
2.0
IP40 in accordance with EN 60529/IEC 60529
For indoor use only
°C
+10 to +30
°C
Guaranteed between +5 and +45
°C
-10 to +60
%
15 to 80 for temperatures up to 30°C non-condensing,
decreasing linearly to 50% relative humidity at 40°C
USB, type B (virtual serial interface)
7-bit ASCII, 1 start bit, 1 or 2 stop bits
Even, odd, none
600 to 57,600 bps
Software or hardware
In accordance with EN 61326-1/IEC61326-1 Electrical equipment for
measurement, control and laboratory use – EMC requirements – Part 1:
General Requirements
Basic requirements
Class B
Suitable for use in residential areas and areas that are connected to a
low voltage network that also supplies residential buildings.
Verified scales in accordance with EU requirements comply with the
requirements of Council Directive 2014/31/EU with EN 45501:2015 and
OIML R76:2006.
* For verified scales in accordance with EU requirements, refer to the
information on the scale.
** For verified scales in accordance with EU requirements, the legal
regulations apply.
Recalculation, formula
Type FSP007-P01P (manufacturer's designation)
100–240 V~, ±10%, 50–60 Hz, ±5%, < 0.2 A
5.2 V
, ± 5%, 1.4 A (max.)
DC
Protection class II
IP40 in accordance with EN 60529/IEC 60529
English
Installation Instructions EVO1S | SPO1S
9
Tabla de contenido
loading

Este manual también es adecuado para:

Pma.hdEvo1sSpo1s