Acelity KCI PREVENA Guía Del Usario página 6

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Indicators and Alerts
Therapy On / Off
Green light shows
therapy is on.
PREVENA™ Incision Dressing Pressure Check
Dressing Not
Compressed?
Leak Alert
one beep, one solid light
Check all connectors
for leaks.
Canister Full Alert
two beeps, one solid light
Check fluid level in canister.
Low Battery Alert
LOW - A slow beep and one solid light -
Be prepared to change batteries.
CRITICAL - A rapid beep and one solid
light - Change batteries right away.
System Error Alert
A repeated beep increasing in volume,
and two solid lights
Hold five seconds
to TURN OFF.
Device Life Cycle Expired
Three lights show life cycle is up. There will be a repeated beep for 15 seconds and then the unit will shut off.
Life cycle = 192 hours (eight days)
Call KCI at 1-800-275-4524.
6
ON / OFF AND AUDIO PAUSE BUTTON
TURN ON - Hold two seconds
TURN OFF - Hold five seconds
AUDIO PAUSE/MUTE - Press once to mute alerts
(See the Audio Pause / Mute Feature section.)
Check for kinked or
pinched tubing.
Make sure canister is securely locked
Press firmly around
and side tabs are flush with unit.
Full or
near full?
Hold five seconds
to TURN OFF.
Hold two seconds
to TURN BACK ON.
Check for leaks.
See Leak Alert
Dressing
section below.
Compressed?
Use PREVENA™ Patch Strips or
dressing edge.
excess KCI Drape to seal leak.
NO
Check for kinked or
pinched tubing.
Turn unit off.
YES
See Therapy On/Off above.
Press and push
Three AA batteries required.
to open panel.
Lithium batteries recommended.
YES
Therapy
running?
NO
Alert Lights
Continue therapy. If alarming for leak,
YES
see the Leak Alert section below.
NO
See Leak Alert section below.
YES
Continue therapy.
Leak Alert
fixed?
Use Audio Pause;
NO
call KCI at 1-800-275-4524
Call KCI at 1-800-275-4524.
See Warnings section.
Press and push
Hold two seconds
to close panel.
to TURN ON.
Continue therapy.
Call KCI at 1-800-275-4524.
Tabla de contenido
loading