799 GPT Titrino
Mode.Parameter.Monitoring.L10Output.Limit
Mode.Parameter.Monitoring.L10Output.Output
Parameters for STAT, DOS, DOC: Output of a signal on L10 output (pin 8) of
the remote socket on limit value violation.
The output can be assigned a definite or all monitoring value(s) as well as the
violated limit and the type of signal to be outputted determined.
3.2.2.51. Mode.Parameter.Presel.RateDisp
Parameters for STAT, DOC: Display of the current rate during the determina-
tion.
3.2.2.52. Mode.Parameter.DosPara.Type
Mode.Parameter.DosPara.Volume.Volume 0.000...10...99999.99
Mode.Parameter.DosPara.Volume.DisType
Mode.Parameter.DosPara.Volume.Rate
Mode.Parameter.DosPara.Volume.Time
Mode.Parameter.DosPara.Time.Time
Mode.Parameter.DosPara.Time.DisType
Mode.Parameter.DosPara.Time.Rate
Mode.Parameter.DosPara.Time.Volume
Mode.Parameter.DosPara.Rate.Rate
Mode.Parameter.DosPara.Rate.StopType
Mode.Parameter.DosPara.Rate.Time
Mode.Parameter.DosPara.Rate.Volume
Parameters for DOS: In the equation Rate = Volume/Time, two parameters
are preset, the third is calculated. Depending on the entry of the type, the fol-
lowing information is required: .Volume, .Time or .Rate.
Entries: .Volume in ml, .Time in s, .Rate in ml/min.
3.2.2.53. Mode.Parameter.DosPara.Pause
Parameters for DOS: Pause time in s.
3.2.2.54. Mode.Parameter.DosPara.TDelta
Parameters for DOS, DOC: Time interval for the entry of a measured point in
the list of measured points.
3.2.2.55. Mode.Parameter.DosPara.Temp
Parameters for DOS, DOC: Temperature in °C. The temperature is measured
continuously and updated if a T sensor is attached.
The temperature is used for temperature compensation in pH measurements.
3.2.2.56. Mode.Parameter.Monitoring.MeasVal.MeasModepH,U,Ipol,Upol,OFF
Mode.Parameter.Monitoring.MeasVal.MeasInput
Mode.Parameter.Monitoring.MeasVal.PCurrent
Mode.Parameter.Monitoring.MeasVal.PVoltage
Mode.Parameter.Monitoring.MeasVal.PolElectrTest
3.2 Remote control commands
meas, temp, rate, all, none
upper,lower,any
active, pulse
volume, time, rate
0.001...150, max.
1...100...999 999
1...100...999 999
volume, rate
0.001...150, max.
0.000...10...99999.99
0.001...150, max.
volume, time
1...100...999 999
0.000...10...99999.99
1...10...999 999
-170.0...25.0...500.0
0...400...±1270
ON, OFF
time, rate
0...999 999
1, 2, diff.
0...1...±127
ON, OFF