799 GPT Titrino
Stop volume in mL = smpl size
The factor is valid.
OFF means that the criterion is not monitored.
3.2.2.21. Mode.Parameter.StopCond.MeasStop
Mode.Parameter.StopCond.UnitMStop
Parameters for DET and MET: Stop when a measured value is reached. Entry
as pH value, in mV (with U and Ipol) and in uA (with Upol). The appropriate
unit can be viewed with .UnitMStop.
OFF means that the criterion is not monitored.
3.2.2.22. Mode.Parameter.StopCond.EPStop
Parameters for DET and MET: Stop when a certain number of EP's has been
found.
OFF means that the criterion is not monitored.
3.2.2.23. Mode.Parameter.StopCond.FillRate
Parameters for DET, MET, SET, KFT, STAT, DOS, DOC: Filling rate in the titra-
tion in mL/min. Max. means maximum possible filling rate with the Exchange
Unit in current use.
3.2.2.24. Mode.Parameter.Statistics.Status
Mode.Parameter.Statistics.MeanN
Mode.Parameter.Statistics.ResTab.Selected
Mode.Parameter.Statistics.ResTab.DelN
Entries for the statistics calculations.
.Status:
On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN:
Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2 Remote control commands
factor
*
pH: 0...±20.00, OFF
U: 0... ±2000, OFF
I:
0...200.0, OFF
read only
1...9, OFF
0.01...150, max.
ON, OFF
2...20
original, delete n,
delete all
1...20
147