Quickstart
Measuring Mode
After the operating voltage has been connected, the analyzer goes
to "Measuring" mode. From the "Calibration" mode, you can switch
the analyzer to "Measuring" mode by pressing the cal key, from the
"Configuration" mode by pressing conf (waiting time for measured-
value stabilization approx. 20 sec).
Sensoface indicator
(sensor status)
Process
variable
Temperature
Mode indicator
(measuring)
Keypad
Sensocheck, Sensoface Sensor Monitoring
Sensocheck continuously monitors the sensor and its wiring.
Sensocheck can be switched on/off (default: off ).
Sensoface provides information on the conductivity sensor
condition. The primary coil and its lines are continuously
monitored for short circuits, the secondary coil and its lines
are checked for open circuits.
The three Sensoface indicators provide information on the
sensor condition.
6