Calibration
Calibration by Input of Cell Factor
Input of cell factor with simultaneous display of conductivity and
temperature (without temperature compensation)
Display
10
Action
Press cal key,
enter code 1100.
Select with key,
edit number with key,
proceed with enter.
Ready for calibration
Remove and clean sensor.
Enter cell factor:
Select using,
enter number using .
Conductivity and temperature
are alternately displayed during
the input (lower display).
Confirm entry with enter.
The entered cell factor and zero
point are displayed.
Confirm with enter.
Remark
Device is in the Hold
mode.
If an invalid code is
entered, the device
returns to measuring
mode.
Display (2 sec)
Device in Hold
mode, measured
value frozen.
Sensoface inactive.