Quickstart
Measuring Mode
After the operating voltage has been connected, the analyzer goes
to "Measuring" mode. From the "Calibration" mode, you can switch
the analyzer to "Measuring" mode by pressing the cal key, from the
"Configuration" mode by pressing conf (waiting time for measured-
value stabilization approx. 20 sec).
Sensoface indicator
(sensor status)
Process
variable
Temperature
Mode indicator
(measuring)
Keypad
Sensocheck, Sensoface Sensor Monitoring
Sensocheck continuously monitors the sensor and its wiring.
Sensocheck can be switched on/off (default: off ).
Sensoface provides information on the sensor condition.
Asymmetry potential, slope, and response time are evaluated
during calibration. The three Sensoface indicators provide
the user with information on wear and required maintenance
of the sensor.
6