Error Messages

Error
Display
ERR 01
Measured
value
blinks
ERR 02
Measured
value
blinks
ERR 98
"Conf"
blinks
ERR 99
"FAIL"
blinks
12
Problem
Possible causes
Sensor
• Wrong cell factor
• Measuring range exceeded
• SAL > 45 ‰
• Sensor connection or cable defective
Unsuitable sensor
Conductance range > 3000 mS
System error
Configuration or calibration data defec-
tive; completely reconfigure the device
using the factory settings. Then calibrate.
Memory error in device program
Factory settings
EEPROM or RAM defective
This error message only occurs in the
case of a total defect. The device must be
repaired and recalibrated at the factory.