CHARACTERISTICS
Size
Weight
Vacuum Pump
Regulator (7)
Gauge (if applicable)
Power Source: Battery (DC Power)
Switch (5)
Collection Canister (13)
Patient tubing
Suction Tip
PROBLEM
Does not function when
switch is on (DC Power)
Blinking LED
Power source connected
indicator light not lit
No suction when pump
is running
General Specifications
S-SCORT VX2
Trouble Shooting
S-SCORT VX2
POSSIBLE CAUSE
Battery discharged
Loose connections
Low battery capacity
Power cord disconnected
Converter defective
Vacuum line (10) loose
Canister (13) defective
Canister lid (13) loose
Catheter thumb vent is open
Debris has collected in pump
8
SPECIFICATIONS
17"L x 9"H x 6"W
Approximately 9 lbs.
12V DC oil-less diaphragm, 3.0A
Exceeds 525 mmHg
Clinical Airflow >= 30LPM
Controls negative pressure
Calibrated in mmHg. Color-coded
Rechargeable Sealed Lead Acid 45 (±10%)
minute running time
Off/Standby/On membrane switch
1200cc SSCOR part #48041
Vinyl tubing 9/32"ID, 72"L SSCOR part
#43200
HI-D
®
. "Big Stick
®
". SSCOR part #44241
CORRECTIVE ACTION
Reconnect to charging source
to activate charger
Tighten connections
Replace battery
Reconnect cord
Check output of converter
Tighten
Replace
Tighten
Occlude
Refer to maintenance
section of this manual