1. Federal law restricts this device to sale, distribution, and use by, or on the order of a physician, emer-
gency medical technician, or other medical practitioner. For use by medical personnel trained in suc-
tioning techniques and in the use of medical suction equipment. To be used only as directed by local
EMS authorities.
2. This manual is restricted to the discussion of the use and maintenance of this device. It does not
attempt to discuss professional techniques in suctioning procedures.
3. Operator should be thoroughly familiar with these operating instructions before this device is used.
4. Use a thumb control on all suction tips, especially when suctioning children and intubated patients.
5. The battery is protected from a deep discharge condition by shutting down after approximately 45
minutes (±10%) running time.
6. Do not use in the presence of flammable agents or anesthetics.
7. The shut-off valve in the canister lid will close down when the canister is full.
8. Install a new canister before testing for vacuum over 300mmHg to minimize the possibility of implo-
sion, which can occur when a canister is aged or damaged.
9. All S-SCORT pumps must be reconnected to the charging source after each use and remain on charge
until needed.