Symbols
Symbols that appear on the display
Symbols
Function/Meaning
Appears while measurement is in progress. It blinks
when the pulse is detected.
Irregular Heartbeat symbol (IHB) Appears when an
irregular heartbeat is detected during measurement
Previous measurements stored in memory
Average data
The battery power indicator during measurement.
The battery is low when it blinks.
Unstable blood pressure due to movement during
E
1
measurement
or
The systolic and diastolic values are within 10
E
mmHg of each other.
2
The pressure value did not increase during inflation.
E
3
The cuff is not applied correctly.
PUL DISPLAY ERROR
E
The pulse is not detected correctly.
Recommended Action
Measurement is in progress. Remain as still as
possible.
––––––––––––––––
––––––––––––––––
––––––––––––––––
––––––––––––––––
Replace all batteries with new ones when the
mark blinks.
Take another measurement. Remain still during
measurement.
Apply the cuff correctly, and take another
measurement.
Apply the cuff correctly, and take another
measurement.