WHAT DISPLAY SYMBOLS MEAN
Display
Symbol
Battery Full
Battery Low
PULSE DISPLAY ERROR
?
E-2
Condition/Cause
The symbol shown while meas-
urement is in progress. It blinks
while detecting the pulse.
An irregular heartbeat or body
movement may have occurred.
Refer to page E-8 for more infor-
mation on irregular heartbeats.
Previous measurement stored in
MEMORY.
The battery power indicator
during measurement.
The batteries are low when it
blinks.
Unstable blood pressure due to
excessive body movement.
The systolic and diastolic values
are within 10 mmHg of each
other.
The pressure value did not
increase during inflation.
The cuff is not fastened correctly.
There is an air leak in the cuff or
monitor.
The pulse is not detected correctly.
Cuff Inflation Meter
Pressure Rating Indicator™
LifeSource Health Line (Toll-Free): 1-888-726-9966
Recommended
Action
Measurement is in progress,
remain as still as possible.
Take measurement again and
consult with your physician.
Replace all batteries with new
ones when the symbol blinks.
Alkaline batteries recommended.
Try the measurement again.
Remain very still during the
measurement.
Fasten the cuff correctly, and try
the measurement again.
Check for air leaks along the
tube and around the
air socket.
Refasten the cuff and retake
measurement.
Make sure tube is properly con-
nected to cuff and monitor.
Try the measurement again.
Remain very still during
the measurement.
Measurement is in progress,
remain as still as possible.
Refer to the section "About
Pressure Rating Indicator" for
further explanation.