Low value inductors (<100uH) are best measured at 10kHz whilst high
values >25H should be measured at 100Hz. The instrument warns if a
measurement is outside its maximum accuracy range by flashing the units
annunciator; if accuracy can be improved by changing the measurement
frequency the frequency annunciator will also flash, see Display section.
Series / Parallel connection
The MZ-805 provides the capability of measuring the series or parallel
equivalent circuit parameters of resistors, capacitors and inductors.
In Auto mode the bridge uses the following models.
These will provide the parameters that will match data sheet values for
most components.
Page 22
Resistor
Inductor
Capacitor <1 µ F
Capacitor >1 µ F
MZ-805 L/C/R meter
Series
Series
Parallel
Series
August 2002
Tabla de contenido