To Use - Beurer FT 95 Manual De Instrucciones

Ocultar thumbs Ver también para FT 95:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

3� Continue in this fashion to set hour, minute, year, month,
day, backlight mode, beep on/off, and Bluetooth
on/off in turn.
4� Press the LIGHT/SET button to save and exit; the device will shut off.
6� To Use
Forehead Temperature Measurement:
NOTES:
• Physical activity, increased perspiration on the forehead, taking vasoconstrictive medication and
skin irritations can distort the measurement.
• The forehead, or the temples, must be free from perspiration and cosmetics.
• If possible, take three measurements on the forehead. If the three measurements differ, use the
highest value.
• If the forehead area is covered in hair, sweat or dirt, clean the area and then wait for 10 minutes
before measurement.
• Keep the thermometer still during measurement. Moving the thermometer during measurement
may cause incorrect results.