Microlife BP3MY1-1P Manual De Instrucciones página 21

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Other possible malfunctions and solutions
If problems occur when using the device, the following points should be
checked and, if necessary, the corresponding measures are to be taken:
Malfunction
The display remains
blank when the instru-
ment is switched on.
The device frequently
fails to measure blood
pressure values, or the
values measured are too
low (too high).
Every measurement pro-
duces a different value.
Blood pressure values
measured at your doctor's
office differ from your
measurements.
Remedy
1. Check batteries for correct polarity.
2. If the display is unusual, re-insert
batteries or exchange them.
1. Check the positioning of the cuff.
2. Measure blood pressure again in peace
and quiet, carefully following the details
in Section 5.
Please read the following information list-
ed in Section 5.2 "Common sources of
error." Repeat the measurement.
Please note: Blood pressure fluctuates
continually so successive measure-
ments will show some variability.
Record the daily development of the mea-
sured values and consult your doctor.
Please note: Individuals visiting their
doctor frequently experience anxiety
which can result in a higher reading
than at home under resting conditions.
19

Capítulos

Tabla de contenido
loading

Tabla de contenido