CONNECTION
INPUT
The discrete I/O interface enables other devices such as PLCs to provide the
triggered voltage (typical 12 VDC or 24 VDC, maximum 30 VDC) to turn the inputs
"on".
An example as the following:
Note:
1.
Voltage: 5~30 VDC, 10 mA maximum current. Voltage polarity cannot be
reversed.
2.
Do not bundle input wiring with power wiring or any other high energy cables.
3.
Wire size: 18 AWG (0.832 mm
24 AWG (0.205 mm
OUTPUT
The four dry-contact normally open relay outputs can be switched up to 250 VAC
or 30 VDC voltages at 1A maximum.
The relay outputs are not polarity-sensitive since they are dry contacts.
An example as the following:
2
) maximum
2
) minimum
EN-5