Acelity PREVENA PLUS Guía Del Paciente página 6

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

PREVENA PLUS™ 125 Therapy Unit
Alert Table
Alert Type
Blockage Alert
Check for blockage:
Full Canister
Leak Alert
Check for leak:
at tubing connectors and
at connection to canister.
Low Battery Alert
System Error Alert
Turn therapy unit off then on.
Off
All LEDs flashing and two
beeps every 15 seconds
6
How to fix
Blocked Tubing
Press firmly around
Use patch strips or
dressing edges.
excess drape to seal leak.
Power Connected
Charging
Fully Charged
On
Continue therapy.
Call KCI at
1-800-275-4524.
Therapy Complete
After seven days of therapy, the PREVENA PLUS™ 125 Therapy Unit will shut off.
Therapy
Closed Clamp
ended alert
Disposal
The therapy unit, power supply, accessories and dressing should be removed by your doctor
or nurse. Do not discard these items with household trash. This could violate local laws
regarding hazardous waste.
Specifications
Transport and Storage Conditions:
Temperature Range: ......................................................................................................0˚F (-18˚C) to 140˚F (60˚C)
Relative Humidity Range: ............................................................................................15 - 95%, non-condensing
Operating Conditions:
Temperature Range: ....................................................................................................... 41˚F (5˚C) to 104˚F (40˚C)
Relative Humidity Range: .............................................................................................15 - 93%, non-condensing
Atmospheric Pressure Range for Optimum Performance: ..................................................................................
The PREVENA PLUS™ Incision Management System is classified as a Type BF applied part
under IEC 60601-1.
IP22 - Protection level against ingress of solid foreign objects and liquids.
All alerts are classified as low priority according to IEC 60601-1-8.
Conforms to: IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-1-8
Electrical Equipment Interference
Cell phones or other similar products could affect the therapy unit. Move the therapy unit
away from these devices if interference is suspected.
Call your doctor or
nurse right away.
1060 hpa (-1253 ft/-381. 9 m) to 700 hpa (9878 ft/3010 m)
Tabla de contenido
loading

Tabla de contenido