Cond Ind One point Slope Calibration
Start as in chapter COND IND ZERO POINT CALIBRATION described. Select the kind of calib r ation.
217.4
µS/cm
25.0
A
°C
Conductivity Calibration
Type = 1 Point Slope
217.4
µS/cm
25.0
A
°C
A Point 1 = 215.0 µS/cm
A C = 217.4 µS/cm
11 Conductivity Quick Setup
For Display setup, analog outputs and setpoints see the section "General Setup".
Sensor Type Selection
1.25
µS/cm
25.0
°C
Sensor Type = Cond(2) m
Cell Constant
1.25
µS/cm
25.0
°C
p M=0.1003 A=0.0000
s M=1.0000 A=0.0000
Measurement units
1.25
µS/cm
25.0
°C
a S/cm
Analog Output? Yes
m
20
Select 1 POINT SLOPE and press [ENTER].
u
Enter the value of calibration point 1 and then press the [ENTER] key to start calibra-
tion.
The cell multiplier and the adder are displayed. After a successful calibration, the
calibration values are taken over and can be stored in the cal history (Adjust), only
u
stored in the cal history (Calibrate) or the calibration can be aborted (Abort).
(PATH: Menu / Quick Setup)
Select the type of sensor to be used with the M400 transmitter. Choices are "Cond(2)",
used for all 2-Electrode type sensors and "Cond (4)" for all 4-electrode sensors.
Press the [ENTER] key.
Enter the appropriate cell constant(s):from the sensor label or certificate (M) for
2-electrode sensors, leaving (A) at 0.000; or (M) and (A) values for 4-electrode
sensors. Press the [ENTER] key.
Select the measurement (conductivity or temperature) and units for measurement.
If using analog output, select YES. Refer back to section "General Setup" to continue
setup.