Sartorius YAS02MI Instrucciones De Instalación página 4

Idiomas disponibles
  • ES

Idiomas disponibles

4 4
§ Loosen the 2 clamp screws (1) and disconnect the
2 power cable leads from the power module.
§ Disconnect the ground wire (earth) (2).
§ Unplug the cable that connects the power module
to the main PCB (3).
§ Open the cable gland and remove the power cord
from the gland.
§ Remove the fastening screws from the power
module (4).
§ Remove the power module from the display and
control unit.
§ Remove the input voltage label ("100 V_240 V")
from the back of the display and control unit
housing.
§ Affix the enclosed voltage label in place of the old
label.
§ Position the 24-V industrial power supply module
inside the display and control unit as shown here.
§ Secure the module with the 2 fastening screws.
loading