Part 2 of 2: High-level disinfecting of transducer for semi-critical
contact
below.
Part 2 of 2: High-level disinfecting of transducer for semi-critical contact
WARNING:
Wear the appropriate personal protective equipment (PPE)
recommended by the chemical manufacturer, such as eyewear and gloves.
High-level disinfectants can cause harm to the patient if not completely
removed from the transducer. Follow the manufacturer's rinse instructions
to remove chemical residue.
Caution:
Do not skip any steps or abbreviate the disinfecting process in any way.
Do not use unapproved disinfecting agents.
Do not soak the transducer longer than recommended by the chemical
manufacturer.
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended disinfectants. Using a non-
recommended disinfecting solution or incorrect solution strength can damage or
discolor the transducer and void the transducer warranty.